Automatic Quantification Software for Geographic Atrophy Associated with Age-Related Macular Degeneration: A
José M Ruiz-Moreno1,2,3, Jorge Ruiz-Medrano2, Francisco Lugo3
1Department of Ophthalmology, Castilla-La Mancha University, Albacete, Spain.
Journal of Ophthalmology
|August 25, 2020
Summary
New software accurately quantifies geographic atrophy (GA) areas in age-related macular degeneration (AMD) using swept-source optical coherence tomography (SS-OCT). The automated tool shows substantial agreement with manual measurements and high repeatability for GA quantification.
Area of Science:
- Ophthalmology
- Medical Imaging
- Computational Biology
Background:
- Age-related macular degeneration (AMD) is a leading cause of vision loss.
- Geographic atrophy (GA) is a key feature of advanced AMD.
- Accurate quantification of GA is crucial for monitoring disease progression and treatment efficacy.
Purpose of the Study:
- To evaluate the accuracy and repeatability of novel software for automated GA area quantification.
- To compare automated measurements with manual assessments using swept-source optical coherence tomography (SS-OCT).
Main Methods:
- Forty-six eyes from 33 AMD patients with GA were scanned using SS-OCT.
- Three masked observers manually measured GA areas from fundus autofluorescence (FAF) images.
- Automated GA area measurements from the software were compared to manual measurements using Lin's concordance correlation coefficient (CCC).
Main Results:
- The overall CCC between manual and automatic GA area measurements was 0.95.
- High agreement was observed for single lesions (CCC=0.99) and regular lesions (CCC=0.97).
- The software demonstrated excellent repeatability with an intraclass correlation coefficient (ICC) of 0.98.
Conclusions:
- The new software provides substantial accuracy and high repeatability for quantifying GA areas in AMD patients.
- The automated tool is particularly precise for measuring single GA lesions.
- This technology has the potential to improve the monitoring of GA progression in clinical practice.