Retinal Vascular Reactivity as Assessed by Optical Coherence Tomography Angiography
Published on: March 26, 2020
Impact of Scan Tilt on Quantitative Assessments Using Optical Coherence Tomography Angiography
Yong Sok Ji1,2,3, Ahmed Roshdy Alagorie1,2,4, Iksoo Byon1,2,5
1Doheny Image Reading Center, Doheny Eye Institute, Los Angeles, CA, USA.
Purpose:
To investigate the impact of scan tilt on quantitative assessments using swept-source optical coherence tomography angiography (SS-OCTA) in healthy participants.
Methods:
Healthy participants were imaged with a SS-OCTA system (PLEX Elite 9000; Carl Zeiss Meditec, Inc, Dublin, CA). After a standard scan was obtained, tilted scans were obtained by moving the optical coherence tomography beam entrance position horizontally. The tilting angle was measured from the B-scan image, and the flattest (horizontal) and the most tilted images were selected for comparative analysis. Foveal avascular zone (FAZ) area, vessel density (VD), and vessel length density (VLD) from the superficial and deep retinal capillary plexus (SCP, DCP), and choriocapillaris flow deficits (CC FDs) were computed and compared between horizontal and tilted images.
Results:
Thirteen eyes were excluded due to poor image quality or small angle difference (<8°) between the horizontal and tilted images. A final cohort of 27 normal eyes of 17 participants with a mean age of 39.3 ± 5.9 years was eligible for analysis. The FAZ area, VD, and VLD of both SCP and DCP were not significantly different between horizontal and tilted images. The CC FD, however, was significantly higher in horizontal images compared with tilted images (21.65% ± 2.41% vs. 21.06% ± 2.19%, P = 0.005).
Conclusions:
CC FD measurements may be significantly affected by the position of the scanning beam and resultant scan tilting. These findings highlight the importance of capturing on-axis (pupil-centered) scans for quantitative OCTA analyses.
Translational Relevance:
By assessing what impact a titled scan can have on OCTA measurements, this study will aid clinicians in understanding how to interpret their results in this situation.
More Related Videos
07:44In vivo Structural Assessments of Ocular Disease in Rodent Models using Optical Coherence Tomography
Published on: July 24, 2020
07:18Evaluation of Capillary and Other Vessel Contribution to Macular Perfusion Density Measured with Optical Coherence Tomography Angiography
Published on: February 18, 2022
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...