Background:
High-frequency (HF) oscillatory ventilation has been shown to improve carbon dioxide (CO2 ) clearance in premature infants. In a previous in vitro lung model with normal lung mechanics we demonstrated significantly improved CO2 washout by HF oscillation of bubble continuous positive airway pressure (BCPAP).
Objective:
To examine CO2 clearance in a premature infant lung model with abnormal lung mechanics via measurement of end-tidal CO2 levels (EtCO2 ) while connected to HF oscillated BCPAP.
Design And Methods:
A 40 mL premature infant lung model with either: normal lung mechanics (NLM): compliance 1.0 mL/cm H2 O, airway resistance 56 cm H2 O/(L/s); or abnormal lung mechanics (ALM): compliance 0.5 mL/cm H2 O, airway resistance 136 cm H2 O/(L/s), was connected to BCPAP with HF oscillation at either 4, 6, 8, 10, or 12 Hz. EtCO2 was measured at BCPAPs of 4, 6, and 8 cm H2 O and respiratory rates (RR) of 40, 60, and 80 breaths/min and 6 mL tidal volume.
Results:
HF oscillation decreased EtCO2 levels at all BCPAPs, RRs, and oscillation frequencies for both lung models. Overall mean ± SD EtCO2 levels decreased (P < .001) from nonoscillated baseline by 19.3 ± 10.2% for NLM vs 14.1 ± 8.8% for ALM. CO2 clearance improved for both lung models (P < .001) as a function of oscillation frequency and RR with greatest effectiveness at 40 to 60 breaths/min and HF at 8 to 12 Hz.
Conclusions:
In this in vitro premature infant lung model, HF oscillation of BCPAP was associated with improved CO2 clearance as compared with nonoscillated BCPAP for both NLM and ALM. The significant improvement in CO2 clearance in an abnormal lung environment is an important step towards clinical testing of this novel respiratory support modality.
Related Concept Videos
Pulmonary Cycle: Exhalation
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Breathing
Acute Respiratory Failure-III