Study Design:
Retrospective radiological analysis.
Objectives:
Translaminar screw (TLS) placement is one of the fixation techniques in the subaxial cervical spine. However, it can be difficult to use in small diameter of the lamina. This study proposed a novel bicortical laminar screw (BLS) and analyzed the related parameters using computed tomography (CT).
Methods:
Cervical CT images taken at our institution from January 2013 to March 2017 were used for measurement. On the axial images, the maximum screw length (MSL) and trajectory angle (TA) of BLS and TLS were measured, together with the distance from the midline (DM) to the BLS entry point and the lamina width (LW). On the parasagittal images, the height of the lamina (LH) was measured.
Results:
MSL of BLS and TLS were 21.00 and 20.97 mm, 19.02 and 20.91 mm, 18.45 and 21.01 mm, and 20.00 and 21.01 mm in C3, C4, C5, and C6, respectively. TA of the BLS and TLS were 21.24° and 34.90°, 19.05° and 34.22°, 18.65° and 33.61°, and 18.30° and 34.51° at C3, C4, C5, and C6, respectively. DM were 6.44, 5.77, 5.68, and 6.03 at C3, C4, C5, and C6, respectively. LW and LH were 3.52 and 12.44 mm, 2.87 and 12.49 mm, 2.76 and 12.42 mm, and 3.18 and 13.30 mm at C3, C4, C5, and C6, respectively.
Conclusion:
We suggest that BLS fixation is a feasible alternative option for posterior fixation to the lamina of the subaxial cervical spine. It may be especially useful when pedicle screw, lateral mass screw, and TLS are not appropriate.