Yasar Sattar1, Waqas Ullah2, Sohaib Roomi2
1Internal Medicine, Icahn School of Medicine at Mount Sinai-Elmhurst Hospital, New York, USA.
Background:
Leadless pacemakers (LPM) are introduced in cardiovascular market with a goal to avoid lead- and pocket-associated complications due to conventional artificial pacemakers (CPM). The comparison of LPM and CPM complications is not well studied at a case by case level.
Methods:
Comprehensive literature was searched on multiple databases performed from inception to December 2019 and revealed 204 cases that received LPM with a comparison of CPM. The data of complications were extracted, screened by independent authors and analyzed using IBM SPSS Statistics for Windows, Version 22.0 (Armonk, NY: IBM Corp.).
Results:
The complications of CPM were high in comparison to LPM in terms of electrode dislodgement (56% vs 7% of cases, p-value < .0001), pocket site infection rate (16% vs 3.4%, p-value = 0.02), and a lead fracture rate (8% vs 0%, p-value = 0.04). LPMs had a statistically non-significant two-times high risk of pericardial effusion (8%) compared to CPMs (4%) with a p-value = 0.8.
Conclusion:
LPMs appear to have a better safety profile than CPMs. There was a low pocket site and lead-related infections in LPM as compared to CPM. However, LPM can have twice the risk of pericardial effusion than CPMs, but this was not statistically significant.