Related Experiment Video
Updated: Dec 10, 2025

A Community-based Stress Management Program: Using Wearable Devices to Assess Whole Body Physiological Responses in Non-laboratory Settings
Published on: January 22, 2018
Stress-Induced Blood Pressure Elevation Self-Measured by a Wearable Watch-Type Device
Naoko Tomitani1, Hiroshi Kanegae1,2, Yuka Suzuki3
1Division of Cardiovascular Medicine, Department of Medicine, Jichi Medical University School of Medicine, Tochigi, Japan.
Background:
Psychological stress contributes to blood pressure (BP) variability, which is a significant and independent risk factor for cardiovascular events. We compared the effectiveness of a recently developed wearable watch-type BP monitoring (WBPM) device and an ambulatory BP monitoring (ABPM) device for detecting ambulatory stress-induced BP elevation in 50 outpatients with 1 or more cardiovascular risk factors.
Methods:
The WBPM and ABPM were both worn on the subject's nondominant arm. ABPM was measured automatically at 30-minute intervals, and each ABPM measurement was followed by a self-measured WBPM measurement. We also collected self-reported information about situational conditions, including the emotional state of subjects at the time of each BP measurement. We analyzed 642 paired BP readings for which the self-reported emotional state in the corresponding diary entry was happy, calm, anxious, or tense.
Results:
In a mixed-effect analysis, there were significant differences between the BP values measured during negative (anxious, tense) and positive (happy, calm) emotions in both the WBPM (systolic BP [SBP]: 9.3 ± 2.1 mm Hg, P < 0.001; diastolic BP [DBP]: 8.4 ± 1.4 mm Hg, P < 0.001) and ABPM (SBP: 10.7 ± 2.1 mm Hg, P < 0.001; DBP: 5.6 ± 1.4 mm Hg, P < 0.001). The absolute BP levels induced by emotional stress self-measured by the WBPM were similar to those automeasured by the ABPM (SBP, WBPM: 141.1 ± 2.7 mm Hg; ABPM: 140.3 ± 2.7 mm Hg; P = 0.724). The subject's location at the BP measurement was also significantly associated with BP elevation.
Conclusions:
The self-measurement by the WBPM could detect BP variability induced by multiple factors, including emotional stress, under ambulatory conditions as accurately as ABPM.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: