Glenoid baseplate screw fixation in reverse shoulder arthroplasty: does locking screw position and orientation
Sejla Abdic1, Jason Lockhart2, Naser Alnusif3
1The Roth|McFarlane Hand and Upper Limb Centre, St. Joseph's Hospital, London, ON, Canada; Department of Surgery, Western University, London, ON, Canada.
Background:
The longevity of a reverse total shoulder arthroplasty prosthesis can be compromised by glenoid baseplate loosening. Circular baseplate designs can be secured with superoinferior (SI) locking screws and anteroposterior (AP) compression screws or vice versa (AP-locking and SI-compression screws). This biomechanical cadaveric study investigated screw position (locking screws SI vs. AP and compression screws AP and SI) and screw orientation (parallel vs. divergent) to determine quantitative differences in baseplate micromotion.
Methods:
Ten paired fresh-frozen cadaveric scapulae (n = 20) were implanted with a standard circular baseplate (∅ = 29 mm). The specimens were randomized into SI-locking or AP-locking screw configurations with the screw orientation directed either parallel or angled divergently at 15°. This yielded a total of 4 groups for statistical comparison: SI-lockingparallel, SI-lockingdivergent, AP-lockingparallel, and AP-lockingdivergent, which were subjected to axial eccentric loading on the implanted baseplates, similar to the American Standard of Testing of Materials standard for shoulder joint arthroplasty.
Results:
In both static and cyclic testing, there were no statistically significant differences (P = .6) in micromotion between SI-locking (2.9 ± 0.8 μm) and AP-locking (3.5 ± 1.5 μm) configurations. In addition, there were no statistically significant differences (P = .2) in the divergent screw orientation group (2.0 ± 0.7 μm) vs. the parallel group (4.0 ± 1.5 μm).
Conclusion:
All configurations of screw position and screw orientation tested in a circular reverse baseplate have similar time-zero fixation in an intact glenoid bone model. In addition, the resultant micromotions for all configurations tested fell far below the 150 μm threshold for bone ongrowth.
More Related Videos
05:57A Teleoperated Robotic System-Assisted Percutaneous Transiliac-Transsacral Screw Fixation Technique
Published on: January 6, 2023
09:51The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve
Published on: September 7, 2022
Related Concept Videos
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Screw: Problem Solving
To evaluate the...
Muscles of the Shoulder
Anterior Thoracic Muscles
The anterior thoracic muscles include the serratus anterior, subclavius, and...