Photon-counting CT with tungsten as contrast medium: Experimental evidence of vessel lumen and plaque visualization
Thomas Sartoretti1, Matthias Eberhard1, Jan Hendrik Rüschoff2
1Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Switzerland.
Background And Aims:
We aimed to investigate the potential of a preclinical photon-counting detector CT (PCT) scanner with an experimental tungsten-based contrast medium for carotid artery imaging.
Methods:
A carotid artery specimen was imaged on a PCT system using the multi-energy bin option (pixel size 0.5 × 0.5 mm2; tube voltage 140 kVp, contrast media-dependent energy thresholds: iodine 20, 75 keV; tungsten 20, 68 keV) at two radiation doses (CTDIvol of 100 mGy and 10 mGy) with iodine and tungsten as contrast media at equal mass-concentrations. Standard CT, virtual non-calcium (VNCa) and calcium-only images were reconstructed. Subjective image quality (4-point Likert scale) was rated using histology as reference. Noise and attenuation measurements were performed. Simulations were conducted to assess the material-decomposition efficiency for different object diameters.
Results:
Image quality on VNCa images was significantly higher for tungsten at lower dose (reader 1/reader 2: 2, [2,2]/2, [2,2] vs 1.5, [2,1]/1, [1,1.75], p < 0.05). Noise was significantly lower at both dose levels for tungsten VNCa images as compared to iodine images (higher dose: tungsten 24 vs iodine 31; lower dose: tungsten 60 vs iodine 82; both p < 0.01). Simulations indicated improved material-decomposition efficiency for tungsten than for iodine pronounced at smaller object diameters.
Conclusions:
PCT employing the multi-energy bin option in combination with tungsten as contrast media enables improved carotid artery imaging with respect to lumen and plaque visualization and image noise.