Usability testing of a smartphone-based retinal camera among first-time users in the primary care setting
Patrick Li1, Yannis M Paulus1,2, Jose R Davila3
1Department of Ophthalmology and Visual Sciences, University of Michigan School of Medicine, Ann Arbor, MI, USA.
BMJ Innovations
|September 1, 2020
Summary
Smartphone retinal photography improves accessibility for diabetic retinopathy screening. Usability testing showed medical assistants quickly mastered the RetinaScope, significantly reducing image acquisition times.
Area of Science:
- Ophthalmology
- Medical Devices
- Primary Care Technology
Background:
- Smartphone retinal photography offers improved accessibility for eye screenings.
- Diabetic retinopathy detection is a key application, but usability challenges persist in primary care.
- The RetinaScope is a novel smartphone-based retinal camera.
Purpose of the Study:
- To assess the usability and efficiency of the RetinaScope for primary care medical assistants with no prior retinal imaging experience.
- To evaluate the learning curve and iterative improvements of the device over multiple testing rounds.
- To determine the feasibility of integrating smartphone retinal photography into routine primary care.
Main Methods:
- Usability testing of the RetinaScope with 24 medical assistants over five iterative rounds.
- Measurement of image acquisition times for single and multiple retinal images using a model eye and human subjects.
- Collection of user feedback through Systems Usability Scale (SUS) scores.
Main Results:
- Significant reduction in image acquisition time for first-time users: single image from 283s to 34s, 5 overlapping images from 325s to 118s (p<0.01).
- In human eye testing, average times were 65s for a single image and 229s for 5 overlapping images.
- High user comfort reported with an average SUS score of 86 ± 13.
Conclusions:
- Smartphone-based retinal photography, exemplified by the RetinaScope, can be rapidly adopted by primary care medical assistants.
- Iterative design improvements enhance device usability and efficiency for non-specialist users.
- This technology holds significant potential for expanding retinal screening accessibility in community settings.