Related Experiment Video
Updated: Dec 10, 2025

Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Multichannel Electrocardiograms Obtained by a Smartwatch for the Diagnosis of ST-Segment Changes
Carmen Anna Maria Spaccarotella1,2, Alberto Polimeni1,2, Serena Migliarino1
1Division of Cardiology, Magna Graecia University, Catanzaro, Italy.
Insights
A smartwatch ECG showed high agreement with standard 12-lead ECG for detecting acute coronary syndromes. This wearable technology offers a feasible and accurate method for diagnosing ST-segment changes, improving patient outcomes.
Area of Science:
- Cardiology
- Medical Devices
- Digital Health
Background:
- Acute coronary syndromes (ACS) are a leading global cause of mortality.
- Timely and accurate diagnosis of ACS is critical for effective intervention and improved patient prognosis.
Purpose of the Study:
- To assess the feasibility and accuracy of a smartwatch in recording multiple electrocardiographic (ECG) leads.
- To compare smartwatch ECG detection of ST-segment changes in ACS with standard 12-lead ECG.
Main Methods:
- A commercially available smartwatch was used to obtain 9-lead ECGs from 100 participants (including ACS patients and healthy individuals).
- Smartwatch ECGs were compared against simultaneous 12-lead ECG recordings.
- Concordance was analyzed using Cohen κ coefficient and Bland-Altman analysis.
Main Results:
- High agreement was observed between smartwatch and standard ECG for normal ECGs (κ=0.90), ST-segment elevation (κ=0.88), and non-ST-segment elevation (κ=0.85).
- Smartwatch ECG demonstrated high sensitivity and specificity for diagnosing normal ECGs, ST elevation, and NSTE ECG alterations.
- Bland-Altman analysis confirmed agreement in detecting ST-segment amplitude.
Conclusions:
- Multichannel smartwatch ECG technology shows significant agreement with standard ECG for identifying ST-segment changes in ACS.
- Smartwatches represent a promising tool for the remote and accessible monitoring of cardiac conditions.
Importance:
Acute coronary syndromes are the leading cause of death worldwide and the leading cause of disease burden in high-income countries. Quick and accurate diagnosis of acute coronary syndromes is essential to avoid fatal events, for timely intervention, and to improve the prognosis.
Objective:
To prospectively investigate the feasibility and accuracy of a smartwatch in recording multiple electrocardiographic (ECG) leads and detecting ST-segment changes associated with acute coronary syndromes compared with a standard 12-lead ECG.
Design, Setting, And Participants:
A commercially available smartwatch was used in 100 participants to obtain multiple-channel ECGs. The study was conducted from April 19, 2019, to January 23, 2020. Fifty-four patients with ST elevation myocardial infarction, 27 patients with non-ST elevation myocardial infarction, and 19 healthy individuals were included in the study. The watch was placed in different body positions to obtain 9 bipolar ECG tracings (corresponding to Einthoven leads I, II, and III and precordial leads V1-V6) that were compared with a simultaneous standard 12-lead ECG.
Main Outcomes And Measures:
The concordance among the results of the smartwatch and standard ECG recordings was assessed using the Cohen κ coefficient and Bland-Altman analysis.
Results:
Of the 100 participants in the study, 67 were men (67%); mean (SD) age was 61 (16) years. Agreement was found between the smartwatch and standard ECG for the identification of a normal ECG (Cohen κ coefficient, 0.90; 95% CI, 0.78-1.00), ST-segment elevation changes (Cohen κ coefficient, 0.88; 95% CI, 0.78-0.97), and non-ST-segment elevation changes (Cohen κ coefficient, 0.85; 95% CI, 0.74-0.96). In addition, the Bland-Altman analysis demonstrated agreement between the smartwatch and standard ECG to detect the amplitude of ST-segment changes (bias, -0.003; SD, 0.18; lower limit, -0.36; and upper limit, 0.36). Use of the smartwatch ECG for the diagnosis of normal ECG showed a sensitivity of 84% (95% CI, 60%-97%) and specificity of 100% (95% CI, 95%-100%); for ST elevation, sensitivity was 93% (95% CI, 82%-99%) and specificity was 95% (95% CI, 85%-99%); and for NSTE ECG alterations, sensitivity was 94% (95% CI, 81%-99%) and specificity was 92% (95% CI, 83%-97%).
Conclusions And Relevance:
The findings of this study suggest agreement between the multichannel smartwatch ECG and standard ECG for the identification of ST-segment changes in patients with acute coronary syndromes.
More Related Videos
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Holter Monitor: 24-Hour Monitoring
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes