Related Experiment Video
Updated: Dec 10, 2025

Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
Catheter navigation support for liver radioembolization guidance: feasibility of structure-driven intensity-based
Houda Hammami1,2, Florent Lalys3, Yan Rolland4
1Therenva, 35000, Rennes, France. houda.hammami@therenva.com.
Purpose:
The fusion of pre/intraoperative images may improve catheter manipulation during radioembolization (RE) interventions by adding relevant information. The objective of this work is to propose and evaluate the performance of a RE guidance strategy relying on structure-driven intensity-based registration between preoperative CTA and intraoperative X-ray images.
Methods:
The navigation strategy is decomposed into three image fusion steps, supporting the catheter navigation from the femoral artery till reaching the injection site (IS). During the pretreatment assessment intervention, the aorta and the origins of its side branches are projected on the intraoperative 2D fluoroscopy following a 3D/2D bone-based registration process, to assist the celiac trunk access. Subsequently, a similar approach consisting in projecting the hepatic vasculature on intraoperative DSA through 3D/2D vessel-based registration is performed to assist the IS location. Lastly, the selected IS is reproduced during the treatment intervention by employing 2D/2D image-based registration between pretreatment and treatment fluoroscopic images.
Results:
The three fusion steps were independently evaluated on subsets of 20, 19 and 5 patient cases, respectively. Best results were obtained with gradient difference as similarity measure and with a delimited preoperative vascular structure for vessel-based registration. The approach resulted in qualitatively appropriate anatomical correspondences when projecting the preoperative structures on intraoperative images. With the best configuration, the registration steps showed accuracy and feasibility in aligning data, with global mean landmarks errors of 1.59 mm, 2.32 mm and 2.17 mm, respectively, a computation time that never exceeded 5 s, 25 s and 11 s, respectively, and a user interaction limited to manual initialization of the 3D/2D registration.
Conclusion:
An image fusion-based approach has been specifically proposed for RE procedures guidance. The catheter manipulation strategy based on the fusion of pre- and intraoperative images has the potential to support different steps of the RE clinical workflow and to guide the overall procedure.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023