D-Dimer Varies Widely Across Instrument Platforms and is Not a Reliable Indicator of Periprosthetic Joint Infections

Lauren N Pearson1, Karen A Moser1, Robert L Schmidt1

  • 1Department of Pathology, ARUP Laboratories & University of Utah, Salt Lake City, UT, USA.

Arthroplasty Today
|September 3, 2020
PubMed

Insights

The D-dimer test shows significant variability across labs, potentially misclassifying patients with periprosthetic joint infection (PJI). A universal cutoff is inappropriate; each lab needs a site-specific D-dimer threshold for accurate PJI diagnosis.

Area of Science:

  • Biomedical diagnostics
  • Clinical chemistry
  • Orthopedic surgery

Background:

  • The D-dimer test is utilized in diagnostic criteria for periprosthetic joint infection (PJI).
  • Standardization issues with D-dimer assays lead to significant inter-laboratory performance variation.
  • This variability impacts the reliability of D-dimer results in clinical decision-making for PJI.

Purpose of the Study:

  • To assess the variability of D-dimer test results across numerous laboratories.
  • To estimate the potential misclassification rate of patients with PJI due to D-dimer assay variability.
  • To evaluate the suitability of a universal D-dimer cutoff for PJI diagnosis.

Main Methods:

  • Analysis of published proficiency testing data from 3903 laboratories.
  • Statistical evaluation of D-dimer result variability.
  • Estimation of patient misclassification rates based on proposed PJI criteria cutoffs.

Main Results:

  • Demonstrated substantial variability in D-dimer results among participating laboratories.
  • Indicated a clinically significant percentage of patients are likely to be misclassified.
  • Highlighted the inappropriateness of a universal D-dimer cutoff for PJI assessment.

Conclusions:

  • The significant inter-laboratory variability of the D-dimer test precludes a universal cutoff for periprosthetic joint infection (PJI) diagnosis.
  • Site-specific studies are essential to establish appropriate D-dimer cutoffs for individual testing platforms.
  • Accurate PJI diagnosis requires addressing D-dimer assay standardization and establishing localized reference ranges.