A forward-viewing radial-array echoendoscope is useful for diagnosing the depth of colorectal neoplasia invasion
Takeshi Kuno1, Takeshi Yamamura2, Masanao Nakamura1
1Department of Gastroenterology and Hepatology, Nagoya University Graduate School of Medicine, Nagoya, Japan.
Background:
It is important to accurately diagnose the depth of colorectal neoplasia invasion. We aimed to evaluate the usefulness of a new forward-viewing radial-array echoendoscope (FRE), relative to the pit pattern method.
Methods:
In this prospective study, the invasion depth of suspected early-stage carcinoma was diagnosed using the pit pattern and FRE methods. The diagnoses were classified as T1a (submucosal invasion distance < 1,000 μm) or shallower and T1b (≥ 1,000 μm) or deeper. Pathological diagnoses were used to compare the diagnostic capabilities of the two methods.
Results:
The final analyses included 110 lesions. The FRE was successfully inserted to the lesion in all cases. There were no significant differences between the two methods regarding the correct prediction rate (79.1% vs. 75.5%, P = 0.57), accuracy (81.3% vs. 79.0%, P = 0.68), specificity (81.3% vs. 70.8%, P = 0.135), positive predictive value (65.0% vs. 60.4%, P = 0.649), negative predictive value (91.0% vs. 98.1%, P = 0.108), or frequency of unevaluable cases (2.7% vs. 4.5%, P = 0.471). The correct prediction rate was calculated by adding the number of unevaluable cases to the denominator of accuracy. Relative to the pit pattern method, the FRE method offered significantly higher sensitivity for predicting T1b or deeper invasion (81.3% vs. 97.0%, P = 0.048). The pit pattern method had significantly poorer accuracy for large lesions (> 36 mm) than for smaller lesions (≤ 36 mm). In contrast, the accuracy of the FRE method did not differ significantly with lesion size. When using the FRE method, no cases were unevaluable because of attenuation. The FRE method provided correct diagnoses in 2 of 3 cases that were unevaluable using the pit pattern method.
Conclusions:
The pit pattern and FRE methods offered similar diagnostic performance for invasion depth. Furthermore, the FRE method may be used to correctly diagnose cases that are unevaluable using the pit pattern method.
More Related Videos
09:42Murine Endoscopy for In Vivo Multimodal Imaging of Carcinogenesis and Assessment of Intestinal Wound Healing and Inflammation
Published on: August 26, 2014
10:35Multimodal Imaging and Spectroscopy Fiber-bundle Microendoscopy Platform for Non-invasive, In Vivo Tissue Analysis
Published on: October 17, 2016
Related Concept Videos
Endoscopic Procedures II: Colonoscopy
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Endoscopic Procedures III: Video Capsule Endoscopy
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Endoscopic Procedures V: ERCP
Patient...