Accuracy of Dual-Energy CT Virtual Unenhanced and Material-Specific Images: A Phantom Study
Baojun Li1, Mark Pomerleau1, Avneesh Gupta1
1Department of Radiology, Boston University School of Medicine, 820 Harrison Ave, FGH BLDG, 3rd Fl, Boston, MA 02118.
Abstract:
OBJECTIVE. The purpose of this study was to determine the quantification accuracy of virtual unenhanced images and establish the lower limit of iodine quantification as a function of dose. MATERIALS AND METHODS. A large elliptical and cylindric phantom mimicking the patient abdomen was scanned on two commercial dual-energy CT scanners, an IQon Spectral CT (Philips Healthcare) and a Revolution CT with Gemstone Spectral Imaging Xtream suite (GE Healthcare). The phantom contained simulated soft tissue, blood, and bone with known elemental composition. It also contained simulated iodine concentrations (0.2-15.0 mg/mL) and iodine-enhanced blood (0.5-5.0 mg/mL). The mean absolute error in CT value for virtual unenhanced images and mean absolute percent error in iodine, calcium, and fat-specific images were measured. RESULTS. For virtual unenhanced images, when excluding the simulated bone, the mean absolute error in CT value was 8.0 ± 5.0 (SD) HU and 9.0 ± 6.2 HU for the IQon and the Revolution CT, respectively (p = 0.61). The mean error in CT value of the simulated bone was -90.5 ± 111.6 HU and -98.5 ± 117.8 HU on the IQon and the Revolution CT, respectively (p = 0.08). For iodine-specific images, the mean absolute percent error was 13.7% and 8.3% for the IQon and the Revolution CT, respectively, above 0.5 mg/mL iodine concentration, and 150% and 100% at less than 0.5 mg/mL iodine concentration. The mean absolute percent error increased from 16.2% at 100% radiation dose to 18.9% and 24% at 75% and 50% dose, respectively, on the IQon; and from 8.8% at 100% dose to 11.1% and 17.8% at 75% and 50%, respectively, on the Revolution CT. CONCLUSION. Virtual unenhanced images are reasonably accurate for simulated soft tissues and contrast materials, except for simulated bone. The lower limit of iodine quantification is radiation-dose dependent. For typical dose levels, 0.5 mg/mL iodine concentration is the lower threshold for iodine detection accuracy.
More Related Videos
07:13Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
08:41Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography