Related Experiment Video
Updated: Dec 10, 2025

Measurement of the Hepatic Venous Pressure Gradient and Transjugular Liver Biopsy
Published on: June 18, 2020
Accurate ultrasonography-based portal pressure assessment in patients with hepatocellular carcinoma
Yu Zhang1, Zhong Wang1, Zhen-Dong Yue1
1Department of Interventional Therapy, Peking University Ninth School of Clinical Medicine, Beijing Shijitan Hospital and Capital Medical University, Beijing 100038, China.
Background:
Portal pressure is of great significance in the treatment of hepatocellular carcinoma (HCC), but direct measurement is complicated and costly; thus, non-invasive measurement methods are urgently needed.
Aim:
To investigate whether ultrasonography (US)-based portal pressure assessment could replace invasive transjugular measurement.
Methods:
A cohort of 102 patients with HCC was selected (mean age: 54 ± 13 years, male/female: 65/37). Pre-operative US parameters were assessed by two independent investigators, and multivariate logistic analysis and linear regression analysis were conducted to develop a predictive formula for the portal pressure gradient (PPG). The estimated PPG predictors were compared with the transjugular PPG measurements. Validation was conducted on another cohort of 20 non-surgical patients.
Results:
The mean PPG was 17.32 ± 1.97 mmHg. Univariate analysis identified the association of the following four parameters with PPG: Spleen volume, portal vein diameter, portal vein velocity (PVV), and portal blood flow (PBF). Multiple linear regression analysis was performed, and the predictive formula using the PVV and PBF was as follows: PPG score = 19.336 - 0.312 × PVV (cm/s) + 0.001 × PBF (mL/min). The PPG score was confirmed to have good accuracy with an area under the curve (AUC) of 0.75 (0.68-0.81) in training patients. The formula was also accurate in the validation patients with an AUC of 0.820 (0.53-0.83).
Conclusion:
The formula based on ultrasonographic Doppler flow parameters shows a significant correlation with invasive PPG and, if further confirmed by prospective validation, may replace the invasive transjugular assessment.
More Related Videos
06:20Contrast-Enhanced Subharmonic Aided Pressure Estimation SHAPE Using Ultrasound Imaging with a Focus on Identifying Portal Hypertension
Published on: December 5, 2020
09:37Invasive Hemodynamic Characterization of the Portal-hypertensive Syndrome in Cirrhotic Rats
Published on: August 1, 2018
Related Concept Videos
Hepatic Portal System
At its core, the hepatic portal vein is the result of a confluence of the superior and inferior mesenteric veins along with the splenic vein. Each of these veins has a unique role. The superior mesenteric vein is...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Effect of Hepatic Disease on Pharmacokinetics: Pathophysiologic Assessment and Liver Function Test