Impact of Patient Alignment on Image Quality in C-Arm Computed Tomography - Evaluation Using an ACR Phantom
Babak Alikhani1, Julius Renne2, Sabine Maschke2
1Center for Radiology and Nuclear Medicine, DIAKOVERE Hospital gGmbH, Hannover, Germany.
Purpose:
To evaluate the influence of patient alignment and thereby heel effect on the image quality (IQ) of C-arm flat-panel detector computed tomography (CACT).
Materials And Methods:
An ACR phantom placed in opposite directions along the z-axis (setup A and B) on the patient support was imaged using CACT. Image acquisition was performed with three different image acquisition protocols. The images were reconstructed with four convolution kernels. IQ was assessed in terms of high contrast using the modulation transfer function (MTF) and low contrast by assessing the image noise, signal-to-noise ratio (SNR) and contrast-to-noise ratios (CNR) as well as the reliability of density measurements. Furthermore, the dose intensity profiles were measured free-in-air.
Results:
The MTF in setup B is higher than the MTF measured in setup A (p < 0.01). The image noises measured in setup A for the air and bone inserts were higher compared to those measured in setup B (p > 0.05). Opposite behavior has been observed for the polyethylene, water-equivalent and acrylic inserts. The SNR for all inserts is inversely related to the image noise. A systematically increasing or decreasing trend of CNR could not be observed (p > 0.05). The intensity profile measured by the detector system free-in-air showed that the anode heel effect is perpendicular to the z-axis.
Conclusion:
The patient alignment has a minor influence on the IQ of CACT. This effect is not based on the X-ray anode heel effect but is caused mainly by the non-symmetrical rotation of CACT.
Key Points:
· The impact of patient alignment and thereby the heel effect on the image quality of CACT was analyzed.. · The patient alignment has a minor influence on the physical parameters related to image quality, such as noise, SNR, and MTF.. · This effect is based mainly on the non-symmetrical rotation of CACT..
Citation Format:
· Alikhani B, Renne J, Maschke S et al. Impact of Patient Alignment on Image Quality in C-Arm Computed Tomography - Evaluation Using an ACR Phantom. Fortschr Röntgenstr 2021; 193: 417 - 426.
More Related Videos
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...