Impact of computed tomography (CT) reconstruction kernels on radiotherapy dose calculation
Irina Vergalasova1, Michael McKenna1, Ning Jeff Yue1
1Department of Radiation Oncology, Rutgers Cancer Institute of New Jersey, Rutgers University, New Brunswick, NJ, USA.
Purpose:
To quantitatively evaluate the effect of computed tomography (CT) reconstruction kernels on various dose calculation algorithms with heterogeneity correction.
Methods:
The gammex electron density (ED) Phantom was scanned with the Siemens PET/CT Biograph20 mCT and reconstructed with twelve different kernel options. Hounsfield unit (HU) vs electron density (ED) curves were generated to compare absolute differences. Scans were repeated under head and pelvis protocols and reconstructed per H40s (head) and B40s (pelvis) kernels. In addition, raw data from a full-body patient scan were also reconstructed using the four B kernels. Per reconstruction, photon (3D and VMAT), electron (18 and 20 MeV) and proton (single field) treatment plans were generated using Varian Eclipse dose calculation algorithms. Photon and electron plans were also simulated to pass through cortical bone vs liver plugs of the phantom for kernel comparison. Treatment field monitor units (MU) and isodose volumes were compared across all scenarios.
Results:
The twelve kernels resulted in minor differences in HU, except at the extreme ends of the density curve with a maximum absolute difference of 55.2 HU. The head and pelvis scans of the phantom resulted in absolute HU differences of up to 49.1 HU for cortical bone and 45.1 HU for lung 300, which is a relative difference of 4.1% and 6.2%, respectively. MU comparisons across photon and proton calculation algorithms for the patient and phantom scans were within 1-2 MU, with a maximum difference of 5.4 MU found for the 20 MeV electron plan. The 20MeV electron plan also displayed maximum differences in isodose volumes of 20.4 cc for V90%.
Conclusion:
Clinically insignificant differences were found among the various kernel generated plans for photon and proton plans calculated on patient and phantom scan data. However, differences in isodose volumes found for higher energy electron plans amongst the kernels may have clinical implications for prescribing dose to an isodose level.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography