Related Experiment Video
Updated: Dec 9, 2025

Retinal Vascular Reactivity as Assessed by Optical Coherence Tomography Angiography
Published on: March 26, 2020
Guidelines for Imaging the Choriocapillaris Using OCT Angiography.
Zhongdi Chu1, Qinqin Zhang1, Giovanni Gregori2
1Department of Bioengineering, University of Washington, Seattle, Washington, USA.
Accurate quantification of choriocapillaris (CC) using optical coherence tomography angiography (OCTA) requires validated visualization and measurement techniques. Researchers must optimize strategies based on instrument, scan patterns, and thresholding for reliable CC parameter analysis.
Area of Science:
- Ophthalmology
- Medical Imaging
- Retinal Imaging
Background:
- Choriocapillaris (CC) plays a crucial role in retinal health.
- Optical coherence tomography angiography (OCTA) is a key imaging modality for CC analysis.
- Standardized methods for CC quantification are needed.
Purpose of the Study:
- To provide guidance on appropriate quantification of choriocapillaris (CC) parameters using optical coherence tomography angiography (OCTA).
- To establish evidence-based recommendations for reliable CC imaging and measurement.
- To address the complexities of quantitative CC analysis in clinical research.
Main Methods:
- Review of existing literature on CC imaging and quantification.
- Analysis of author experiences with OCTA in retinal research.
- Evaluation of visualization and measurement validation criteria for CC parameters.
Main Results:
- Accurate CC visualization requires physiologically sound slabs and consistent imaging across scans.
- Validated CC quantification depends on physiologically meaningful parameters and appropriate binarization algorithms.
- Intercapillary distances (ICDs) measured by OCTA should align with histology and adaptive optics data.
- Optimization of thresholding methods, like Phansalkar, is crucial, with window radius related to average ICD.
Conclusions:
- Quantitative CC analysis with OCTA is complex and instrument-dependent.
- Researchers must tailor OCTA strategies to specific instruments, scan patterns, anatomy, and thresholding techniques.
- Achieving accurate and reliable CC measurements necessitates careful consideration of multiple factors.