Variability of computed tomography radiomics features of fibrosing interstitial lung disease: A test-retest study
Florian Prayer1, Johannes Hofmanninger2, Michael Weber1
1Department of Biomedical Imaging and Image-guided Therapy, Medical University of Vienna, Vienna, Austria.
Objectives:
To investigate the intra- and inter-scanner repeatability and reproducibility of CT radiomics features (RF) of fibrosing interstitial lung disease (fILD).
Methods:
For this prospective, IRB-approved test-retest study, CT data of sixty fILD patients were acquired. Group A (n = 30) underwent one repeated CT scan on a single scanner. Group B (n = 30) was scanned using two different CT scanners. All CT data were reconstructed using different reconstruction kernels (soft, intermediate, sharp) and slice thicknesses (one and three millimeters), resulting in twelve datasets per patient. Following ROI placement in fibrotic lung tissue, 86 RF were extracted. Intra- and inter-scanner RF repeatability and reproducibility were assessed by calculating intraclass correlation coefficients (ICCs) for corresponding kernels and slice thicknesses, and between lung-specific and non-lung-specific reconstruction parameters. Furthermore, test-retest lung volumes were compared.
Results:
Test-retest demonstrated a majority of RF is highly repeatable for all reconstruction parameter combinations. Intra-scanner reproducibility was negatively affected by reconstruction kernel changes, and further reduced by slice thickness alterations. Inter-scanner reproducibility was highly variable, reconstruction parameter-specific, and greatest if either soft kernels and three-millimeter slice thickness, or lung-specific reconstruction parameters were used for both scans. Test-retest lung volumes showed no significant difference.
Conclusion:
CT RF of fILD are highly repeatable for constant reconstruction parameters in a single scanner. Intra- and inter-scanner reproducibility are severely impacted by alterations in slice thickness more than reconstruction kernel, and are reconstruction parameter-specific. These findings may facilitate CT data and RF selection and assessment in future fILD radiomics studies collecting data across scanners.
More Related Videos
10:17Guidelines and Experience Using Imaging Biomarker Explorer IBEX for Radiomics
Published on: January 8, 2018
07:38A Multimodal Imaging Approach Based on Micro-CT and Fluorescence Molecular Tomography for Longitudinal Assessment of Bleomycin-Induced Lung Fibrosis in Mice
Published on: April 13, 2018
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...