Related Experiment Video
Updated: Dec 9, 2025

A Spine Robotic-Assisted Navigation System for Pedicle Screw Placement
Published on: May 11, 2020
3D-imaging in percutaneous spine surgery using the Surgivisio system
Mehdi Boudissa1, Marc Prod'homme2, Gaël Kerschbaumer2
1Service de chirurgie orthopédique et traumatologique, CHU Grenoble, hôpital Nord, boulevard de la Chantourne, 38700 La Tronche, France; Université Grenoble Alpes, Grenoble, France; CNRS UMR 5525, laboratoire TIMC-IMAG, université Grenoble Alpes, Pavillon Taillefer, 38700 La Tronche, France.
Abstract:
The Surgivisio system is a new generation of intraoperative 3D imaging, used in our institution for pedicle insertion in minimally invasive spine surgery since January 2018. The purpose of this technical note is to describe the Surgivisio system, its advantages and its limitations, in percutaneous spinal surgery. Results of the first year of use were analyzed, in a series of 29 patients, to evaluate accuracy of pedicle screw insertion, operative time and radiation exposure. On the Heary and Gertzbein classifications, 95.5% of pedicle screw placements (107/112) were rated as acceptable. Mean operative time was 29.3min per vertebra; mean radiation exposure per vertebra was 0.61 mSv. The Surgivisio system is an effective navigation tool for pedicle screw insertion in minimally invasive spinal surgery, with acceptable radiation exposure and operative time for each navigated vertebra. LEVEL OF EVIDENCE: II, prospective cohort study.