Optimum waveform envelopes and amplitude ratios in oscillometric blood pressure estimation
Mariana Alvarado Alvarez1, Raj Padwal2, Jennifer Ringrose2
1Medicine, University of Alberta.
Objective:
To determine if, when using the oscillometric method, there is a specific range of amplitude ratios in the fixed-ratio algorithm that will result in blood pressure estimates that consistently fall within a mean error ≤5 mmHg and a SD of the error <8 mmHg. Additionally, to apply different representations of the oscillometric waveform envelope to verify if this will affect the accuracy of the results.
Methods:
SBP and DBP were obtained using the fixed-ratios method applied to a dataset of 219 oscillometric measurements obtained from 73 healthy volunteers and compared to their corresponding auscultation values. Ratio and envelope analysis were done on Matlab (The MathWorks, Inc., Natick, Massachusetts, USA).
Results:
Depending on the envelope representation, ratios between 0.44-0.74 for systolic pressure and 0.51-0.85 for diastolic pressure yield results within the limits mentioned above. When a set of optimum envelope representations and ratios are selected based on population mean, the highest percentage of subjects presenting blood pressure estimates within the limits were 72.6% for systolic and 69.9% for diastolic.
Conclusion:
The range of ratios presenting optimum results appears to be independent of the degree of arterial stiffness given the wide range of ages of the subjects in the study. Different representations of the oscillometric waveform envelope may improve the accuracy of the method. However, there remains a considerable percentage of the population with unreliable results. It is therefore important to only use devices that have been properly validated according to standard protocol.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)