Related Experiment Video
Updated: Dec 9, 2025

3D Ultrasound Imaging: Fast and Cost-effective Morphometry of Musculoskeletal Tissue
Published on: November 27, 2017
A Gel Pad Designed to Measure Muscle Volume Using Freehand 3-Dimensional Ultrasonography
Julien Thomare1, Lilian Lacourpaille1, Peter J McNair2
1Université de Nantes, Movement, Interactions, and Performance Laboratory, Nantes, France.
Abstract:
We developed an innovative gel pad that covers the entire lower leg to remove artifacts due to the pressure of the transducer in freehand 3-dimensional ultrasonography. In comparison to the reference method in water, this study showed that this new method was valid (bias, 3.4 mL; limit of agreement, 7.7 mL for a volume of ≈220 mL) and reliable (coefficient of variation, <1.1%) for the measurement of gastrocnemius medialis muscle volume. Considering that it is easier to use than the water tank technique, it has much promise for volumetric measurement of many muscles.
More Related Videos
09:57Obtaining Quality Extended Field-of-View Ultrasound Images of Skeletal Muscle to Measure Muscle Fascicle Length
Published on: December 14, 2020
09:21Muscle Function Obtained with Motion Mode Ultrasound and Surface Electromyography during Core Endurance Exercise
Published on: August 25, 2022