Related Experiment Video
Updated: Dec 9, 2025

Proximal Cadaveric Femur Preparation for Fracture Strength Testing and Quantitative CT-based Finite Element Analysis
Published on: March 11, 2017
Early Failure of a Carbon-Fiber-Reinforced Polyetheretherketone Distal Femur Plate: A Case Report
W Alexander Goudriaan1, Rutger L Tordoir, Demien Broekhuis
1Department of Orthopaedic Surgery, Leiden University Medical Center- Leiden, the Netherlands.
Case:
In this case, we report on a carbon-fiber-reinforced polyetheretherketone plate failure 4 months after implantation, radiation therapy, chemotherapy, and protective weight-bearing in a 75-year-old woman who sustained a nontraumatic pathological distal femur fracture due to lymphoma.
Conclusion:
Although carbon-fiber composite implants are regularly used and, to date, there have been no reports of early clinical failures, the longevity of the implant's structural integrity after high-dose radiation and/or chemotherapy treatment has not been fully explored. Therefore, we deem it too early to conclude that carbon-fiber-reinforced polyetheretherketone implants are superior to conventional implants in treating (pathological) fractures.
More Related Videos
06:38Establishment of a Segmental Femoral Critical-size Defect Model in Mice Stabilized by Plate Osteosynthesis
Published on: October 12, 2016
08:20A Reliable and Reproducible Critical-Sized Segmental Femoral Defect Model in Rats Stabilized with a Custom External Fixator
Published on: March 24, 2019