Translational Rabbit Model of Chronic Cardiac Pacing
Published on: January 6, 2023
Assessment of injury current during leadless pacemaker implantation
Alexander Breitenstein1, Ardan M Saguner1, Alessio Gasperetti1
1Electrophysiology, Department of Cardiology, University Heart Center, University Hospital Zurich, Zurich, Switzerland.
Background:
Leadless pacemakers are an established treatment option for bradyarrhythmias. Similar to conventional transvenous pacemakers, satisfying pacing values during implantation are targeted for optimal long-term device function. The objective is to investigate the role of a local injury current (IC) in leadless pacemaker implantations.
Method:
The IC, sensing value, capture threshold and impedance were collected in 30 consecutive patients receiving a leadless pacemaker.
Results:
39 EGMs were recorded from 30 patients (including 9 device repositions). An IC was detected in 15 cases (38%). At implantation, the presence of an IC was associated with a significantly lower sensing (7.1 ± 3.7 mV vs 12.0 ± 4.0 mV; P = 0.004) and a higher capture threshold (median threshold 1.13 V at 0.24 ms [0.50-2.00] vs 0.50 V at 0.24 ms [0.25-0.75]; P = 0.002) and with a 26 fold higher likelihood of device repositioning compared to the absence of an IC (OR 26.3 [2.79-248], P < 0.001). Patients with an IC in their final implant position had a lower sensing (9.3 ± 4.4 mV vs 13.6 ± 4.7 mV at implantation, P = 0.04), while the initially similar capture threshold was lower after 24 h in the IC group. After 2 weeks, all parameters were similar between the two groups.
Conclusions:
Our study shows that an IC can readily be observed during leadless pacemaker implantation associated with a lower sensing and a higher capture threshold at implantation but with similar to even better values during follow-up.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Cardiac Catheterization IV: Nursing Management
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...