Three Different Protocols of Corneal Collagen Crosslinking in Keratoconus: Conventional, Accelerated and Iontophoresis
Published on: November 12, 2015
Clinical utility of irx3 in keratoconus
Maria Markoulli1, Cathleen Fedtke1, Minas Coroneo2
1School of Optometry and Vision Science, The University of New South Wales , Sydney, Australia.
Clinical Relevance:
Diagnosis and monitoring of keratoconus is increasingly being conducted with the aid of imaging equipment such as corneal aberrometry. There is a need to also know the confidence with which ocular aberration measurements can be made.
Background:
To assess the repeatability of lower- and higher-order aberration measurements in patients with keratoconus using the irx3 wavefront aberrometer (Imagine Eyes, Orsay, France) and evaluate correlations with corneal curvature.
Methods:
The irx3 wavefront aberrometer was used to measure bilateral lower- and higher-order ocular aberrations on 33 participants with keratoconus. Three measurements were taken from each eye to determine the repeatability of lower-order aberrations (quantified as sphere and cylinder in dioptres) and higher-order aberration co-efficients (up to eighth order in micrometres), coma, trefoil and total higher-order aberration root mean square (in micrometres). Corneal curvature was measured using the Pentacam HR system (OCULUS, Wetzlar, Germany).
Results:
Repeat measurements for lower-order aberrations resulted in larger co-efficients of repeatability than higher-order aberrations. Similarly, larger co-efficients of repeatability between repeated measures across all Zernike co-efficients were observed in eyes with severe keratoconus (that is, corneal curvature > 52-D) compared to eyes with flatter corneas. The difference between repeated measures tended to be significant for the lower-order aberrations regardless of corneal curvature. The highest correlations with corneal curvature for right and left eyes respectively, were identified for total higher-order aberration root mean square (r = 0.92, p < 0.001 and r = 0.91, p < 0.001), followed closely by coma (r = -0.93, p < 0.001 and r = -0.86, p < 0.001) and the Z (3, -1) co-efficient (r = -0.92, p < 0.001 and r = -0.86, p < 0.001 for right and left eyes, respectively).
Conclusions:
Lower-order aberrations tended to be less repeatable, indicating that instrument variability must be considered when monitoring progression. Total higher-order aberration root mean square and third-order aberrations, in particular the vertical coma Z (3, -1) co-efficient, demonstrated a stronger correlation with corneal curvature than the lower-order aberrations.