Related Experiment Video
Updated: Jul 26, 2026

Electronic Tongue Generating Continuous Recognition Patterns for Protein Analysis
Published on: September 16, 2014
Robotic sheath-flow probe electrospray ionization/mass spectrometry (sfPESI/MS): development of a touch sensor for
Kenzo Hiraoka1, Osamu Ariyada2, Ryo Sekine3
1Clean Energy Research Center, University of Yamanashi, 4-3-11, Takeda, Kofu, Yamanashi 400-8511, Japan. hiraoka@yamanashi.ac.jp.
Abstract:
In the previous work, sheath-flow probe electrospray ionization (sfPESI) equipped with a touch sensor was developed for conducting samples. In this work, a capacitiance-sensitive touch sensor that can be applicable to samples prepared in a nonconducting plastic multiwell plate was developed. The radiofrequency with 5 kHz and 4.5 Vpp was applied to the metal substrate on which the plastic plate was placed. The probe tip stopped at the position where it touched the surface of the liquid solution prepared in the plastic multiwell plate by detecting the displacement current flowing through the capacitance of the circuit. By coupling a nondisposable sfPESI probe with a table-top 3-axis robot, consecutive analysis of peptides, proteins, drugs, and real samples was performed. The carry-over by the consecutive analyses was suppressed to minimal by cleansing the probe tip using the solvent of water/methanol/acetonitrile (1/1/1).
More Related Videos
15:00Applications of the Single-probe: Mass Spectrometry Imaging and Single Cell Analysis under Ambient Conditions
Published on: June 14, 2016
07:55Integrated Cell Manipulation Platform Coupled with the Single-probe for Mass Spectrometry Analysis of Drugs and Metabolites in Single Suspension Cells
Published on: June 21, 2019