X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Dose-reducing fluoroscopic system decreases patient but not occupational radiation exposure in chronic total
Pablo Salinas1, Roberto M Sanchez-Casanueva2,3, Nieves Gonzalo1
1Cardiology Department, Hospital Clínico San Carlos and Instituto de Investigación Sanitaria del Hospital Clínico San Carlos (IdISSC), Madrid, Spain.
Aims:
Several novel low-dose fluoroscopic systems (LDS) developed recently, but real practice information of the net benefit for the patient and professionals is scarce. We evaluated separately patient and operator radiation exposure during percutaneous interventions of chronic total occlusions (CTO).
Methods:
A total of 116 consecutive CTOs were analyzed (60 in LDS and 56 in standard-dose fluoroscopic system [SDS]). Digital dosimetry of patient and occupational (operator and scatter dose) exposure was prospectively recorded.
Results:
Biometrics, demographics, CTO variables, and operators were distributed evenly. Patient radiation exposure was effectively decreased in LDS (dose area product [DAP] by 36%, Air Kerma [AK] by 47%). However, occupational data showed no statistical differences between LDS and SDS. The LDS uses less radiation amount but with higher energy (due to additional filtration) compared to SDS, therefore increasing the scatter dose. When comparing the C-arm scatter dose to the DAP we found higher scatter dose with the LDS (0.0139 mSv/gray (Gy)*cm2 vs. 0.0082 mSv/Gy*cm2, p < .001). This was confirmed in a larger dataset comprising 5,221 coronary procedures.
Conclusions:
LDS was safer for patients reducing DAP and AK compared to SDS. However, occupational doses were not lower and scatter dose higher. Radiological protection measures must be kept maximized even in LDS.
More Related Videos
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Dosage Regimen: Fixed Dose
Fixed-dose regimens can be used for various routes of administration, including intravenous (IV) injections and oral medications. For IV administration, a predetermined amount of the drug is...