3D cone-beam tomosynthesis provides axial imaging of the spine with lower radiation compared to computed tomography
Vidyadhar V Upasani1,2, Harsha Bandaralage3, Christine L Farnsworth3
1University of California San Diego, San Diego, CA, USA. vupasani@rchsd.org.
Study Design:
Three imaging techniques were compared using porcine spines.
Objectives:
To compare image acquisition time, radiation exposure, pedicle width measurement, assessment of screw breach, and image artifact between cone-beam tomosynthesis (CBT) single mode, CBT dual mode (stereotactic CBT), and computed tomography (CT) imaging with and without spinal implants. CT is the standard for axial imaging of orthopedic procedures. CBT technology is being developed, allowing real-time intraoperative imaging and 3D surgical guidance. CBT may deliver useful axial imaging quicker with less radiation than current technologies.
Methods:
Six porcine spines were instrumented with bilateral pedicle screws at six levels connected with 5.5 mm rods. Dosimeters were attached to four surfaces of spines. CT, CBT single and CBT dual images were acquired pre-implant and post-implant. Image acquisition and 3D reconstruction times were recorded. Pedicle widths were measured before and after instrumentation. Screw medial breaches were graded (0: no breach, 1: < 2 mm, 2: 2-4 mm, 3: > 4 mm). Artifact and/or distortion of each image was ranked (0 = none, 1 = mild, 2 = moderate, 3 = large). Image acquisition and reconstruction times, radiation dose, pedicle width, screw breach and artifact were compared between techniques.
Results:
Total image acquisition and reconstruction times of CBT was significantly less (single: 9.9 ± 0.2 s, p < 0.001; dual: 60.0 ± 8.7 s, p < 0.001) than CT (250.3 ± 36.7 s). CBT had significantly less radiation exposure than CT (CT: 0.7 ± 0.1 rad, single: 0.03 ± 0.02 rad, dual: 0.07 ± 0.03 rad; p < 0.001). No difference in pedicle width change pre-implant to post-implant was found (CT: p = 0.449, single: p = 0.430, dual: p = 0.528). Pedicle width (pre-implant: p > 0.5, post-implant: p > 0.9) and pedicle width change (p > 0.4) was similar amongst all techniques. Breach assessment was not different between groups (p = 0.257). CBT images had consistently lower artifact grades than CT.
Conclusions:
Although CBT axial image quality appeared subjectively inferior to CT, it enabled consistent assessment of pedicle width and screw breach, at half time and 10× lower radiation exposure. With continued refinements, CBT technology may allow for adequate intra-operative axial imaging using low radiation exposure.
More Related Videos
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
08:34Optimizing Minimally Invasive Spine Surgery: A Fully 3D CT O-Arm Navigated Workflow in MIS TLIF
Published on: October 17, 2025
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
X-ray Imaging
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Radiological Investigation I: X-ray and CT