Performance Characteristics of Positive Expiratory Pressure Devices
Angela M Demchuk1, Robert L Chatburn2
1Cleveland Clinic, Cleveland, Ohio.
Background:
Positive expiratory pressure (PEP) therapy imposes expiratory flow resistance to increase airway diameter and enhance mucus clearance. PEP is achieved several ways. Oscillatory PEP devices (OPEP) generate repeated occlusions that are known to reduce mucus viscosity. There are many marketed devices, but comparative performance is mostly unreported. The purpose of this study was to evaluate performance characteristics of many PEP/OPEP devices. For OPEP devices, we defined an optimal performance metric by creating an oscillation index that combines the OPEP performance characteristics.
Methods:
PEP devices (TheraPEP, EzPAP, VersaPAP, Resistex, AccuPEP, AccuPAP, and Threshold PEP) and OPEP devices (Acapella DH, Acapella DM, Acapella Choice, ShurClear, Aerobika, VibraPEP, vPEP, and PocketPEP with and without the Oxyjet attachment) were tested by adjusting simulated expiratory flow from 5 L/min to 30 L/min in increments of 5 L/min using a standard flow meter.
Results:
All devices showed varying performance characteristics. As expiratory flow increased, mean PEP increased for most devices. The TheraPEP showed a mean PEP of 13 cm H2O across all settings. For OPEP devices, there was a major difference between pressure and flow waveforms. The Acapella DH, ShurClear, and Aerobika showed the highest flow amplitude, flow frequency, and oscillation index.
Conclusions:
PEP devices behaved similarly and as expected, with increased pressure with increased flow (flow resistors) or flow independence (threshold resistors). There was much greater variation in the performance of the OPEP devices. A higher oscillation index indicates better mechanical performance characteristics. Many devices have similar characteristics. However, the devices with the highest oscillation index have the highest flow amplitude and frequency, which may indicate better clinical performance.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Pulmonary Cycle: Exhalation
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...