Evaluation of Image Quality for 7 Iterative Reconstruction Algorithms in Chest Computed Tomography Imaging: A Phantom Study
Related Experiment Video
Updated: Dec 8, 2025

Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Kristin Jensen, Guro Hagemo1, Anders Tingberg2
1Department of Radiology and Nuclear Medicine, Radiumhospitalet, Oslo University Hospital, Oslo, Norway.
Objectives:
This study aimed to evaluate the image quality of 7 iterative reconstruction (IR) algorithms in comparison to filtered back-projection (FBP) algorithm.
Methods:
An anthropomorphic chest phantom was scanned on 4 computed tomography scanners and reconstructed with FBP and IR algorithms. Image quality of anatomical details-large/medium-sized pulmonary vessels, small pulmonary vessels, thoracic wall, and small and large lesions-was scored. Furthermore, general impression of noise, image contrast, and artifacts were evaluated. Visual grading regression was used to analyze the data. Standard deviations were measured, and the noise power spectrum was calculated.
Results:
Iterative reconstruction algorithms showed significantly better results when compared with FBP for these criteria (regression coefficients/P values in parentheses): vessels (FIRST: -1.8/0.05, AIDR Enhanced: <-2.3/0.01, Veo: <-0.1/0.03, ADMIRE: <-2.1/0.04), lesions (FIRST: <-2.6/0.01, AIDR Enhanced: <-1.9/0.03, IMR1: <-2.7/0.01, Veo: <-2.4/0.02, ADMIRE: -2.3/0.02), image noise (FIRST: <-3.2/0.004, AIDR Enhanced: <-3.5/0.002, IMR1: <-6.1/0.001, iDose: <-2.3/0.02, Veo: <-3.4/0.002, ADMIRE: <-3.5/0.02), image contrast (FIRST: -2.3/0.01, AIDR Enhanced: -2.5/0.01, IMR1: -3.7/0.001, iDose: -2.1/0.02), and artifacts (FIRST: <-3.8/0.004, AIDR Enhanced: <-2.7/0.02, IMR1: <-2.6/0.02, iDose: -2.1/0.04, Veo: -2.6/0.02). The iDose algorithm was the only IR algorithm that maintained the noise frequencies.
Conclusions:
Iterative reconstruction algorithms performed differently on all evaluated criteria, showing the importance of careful implementation of algorithms for diagnostic purposes.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...