High-sensitivity flow cytometric assays: Considerations for design control and analytical validation for
Ulrike Sommer1, Steven Eck2, Laura Marszalek3
1Novartis Institutes of Biomedical Research, Novartis Pharma AG, Basel, Switzerland.
Abstract:
The current consensus recommendation papers dealing with the unique requirements for the analytical validation of assays performed by flow cytometry address the validation of sensitivity (both analytical and functional) only in general terms. In this paper, a detailed approach for designing and validating the sensitivity of rare event methods is described. The impact of panel design and optimization on the lower limit of quantification (LLOQ) and suggestions for reporting data near, or below, the LLOQ are addressed. This paper serves to provide best practices for the development, optimization, and analytical validation of flow cytometric assays designed to assess rare events. Note that this paper does not discuss clinical sensitivity validation, which addresses the positive and negative predictive value of the test result.