Can different Catphan phantoms be used in a multi-centre audit of radiotherapy CT image quality?
Anne T Davis1, Antony L Palmer1, Andrew Nisbet2
1Department of Medical Physics and Biomedical Engineering, University College London, UK; Department of Medical Physics, Portsmouth Hospitals NHS Trust, Portsmouth, UK.
Purpose:
To determine the variation between Catphan image quality CT phantoms, specifically for use in a future multi-centre image quality audit.
Method:
14 Catphan phantoms (models 503, 504 and 604) were scanned on a Canon Aquilion Prime CT scanner using a single scan protocol. Measurements were made of noise in the uniformity section, visibility of low contrast targets and contrast, x-ray attenuation and CT number for 5 materials in the sensitometry section. Scans were also acquired using one phantom and varying reconstruction field of view, image slice thickness, effective tube-current-time product and iterative reconstruction settings to determine how the degree of inter-phantom variability compared with the magnitude of changes from scan parameter alteration.
Results:
Across all phantoms the mean CT value in the uniformity section was 7.0 (SD 0.9) range: 4.9-8.1 HU. For the different materials the CT numbers were air: -1004 ± 5, Polymethylpentene: -190 ± 2, Polystyrene: -42 ± 2, Delrin: 321 ± 5 and Teflon: 898 ± 8 HU. Consistency of low contrast targets through visual scoring was good. Measured contrast was lower (p < 0.001) with more variability for 504 versus 604 models. All phantoms produced identical tube current settings with x-ray tube current modulation, indicating no x-ray attenuation differences. The degree of change in image quality metrics between phantoms was small compared with results when scan parameters were varied.
Conclusion:
Catphan phantoms model 604 showed minimal differences and will be used for multi-centre inter-comparison work, with the consistency between phantoms appropriate for measuring possible variations in image quality.