Flexor Pollicis Longus Tendon Wear Associated With Volar Plating: A Cadaveric Study
Nicole A Zelenski1, Elaine Schmidt2, Matthew Chin2
1Department of Orthopaedic Surgery, University of Pennsylvania, Philadelphia, PA.
The Journal of Hand Surgery
|September 20, 2020
Summary
Low-profile volar rim plates and distally placed plates increase flexor pollicis longus tendon contact pressure and wear compared to proximally placed plates in a cadaver model. Mechanical properties remained largely unchanged across all plate types.
Area of Science:
- Orthopedic surgery
- Biomechanical engineering
- Hand surgery
Background:
- Distal radius fractures often require surgical fixation.
- Volar locking plates are commonly used, but plate prominence can affect adjacent tendons.
- The flexor pollicis longus (FPL) tendon is at risk of iatrogenic injury from volar implants.
Purpose of the Study:
- To evaluate the effect of different volar plate designs on the flexor pollicis longus (FPL) tendon.
- To compare contact pressures, wear patterns, and mechanical properties of the FPL tendon with low-profile volar rim (VR) plates, proximally placed variable-angle locking plates (pVA-LCP), and distally placed variable-angle locking plates (dVA-LCP).
Main Methods:
- Cadaveric specimens (9 pairs) were used to assess plate-tendon interactions.
- Thin-film pressure sensors measured initial contact loads.
- Cyclic loading (10,000 cycles) simulated tendon actuation.
- Tendon wear was graded using stereomicroscopy and a blinded Likert scale.
- Uniaxial tensile testing evaluated tendon mechanical properties.
Main Results:
- Both VR plates and dVA-LCP resulted in increased contact pressure and tendon wear compared to pVA-LCP.
- No significant difference in contact pressure or wear was observed between VR and dVA-LCP groups.
- No significant differences in ultimate failure load, stress, stress relaxation, elastic modulus, or stiffness were found among the groups.
Conclusions:
- Plates positioned on or beyond the volar rim lead to elevated contact pressures and increased tendon wear.
- The proximally placed variable-angle locking plate (pVA-LCP) demonstrated the lowest contact pressure and wear.
- While low-profile plates aid fixation, their potential for FPL tendon irritation necessitates careful consideration and possible removal.