Related Experiment Video
Updated: Dec 8, 2025

Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025
Radiation Exposure During Transcatheter Aortic Valve Replacement: Impact of Arterial Approach and Prosthesis Type
Laurent Faroux1, Aurélien Villecourt2, Leonardo Guimaraes3
1Quebec Heart and Lung Institute, Laval University, Quebec City, Quebec, Canada; Department of Cardiology, Reims University Hospital, Reims, France; EA3797 VieFra, University of Reims Champagne-Ardenne, Reims, France.
Background:
The impact of novel alternative access and valve type on radiation exposure during transcatheter aortic valve replacement (TAVR) has not yet been evaluated. This study sought to determine the impact of a transarterial approach and prosthesis type on physician and patient exposure to radiation during TAVR.
Methods:
This was a prospective study including 140 consecutive patients undergoing TAVR by transfemoral (n = 102) or transcarotid (TC) (n = 38) access at 2 centers. Implanted valves were the self-expanding Evolut R/PRO system (Medtronic, Minneapolis, MN; n = 38) and the balloon-expandable SAPIEN 3 valve (Edwards Lifesciences, Irvine, CA; n = 102). The primary endpoint was first operator radiation exposure. The secondary endpoint was patient radiation exposure.
Results:
First operator radiation exposure was 4-fold greater during TC TAVR (P < .001). The use of a self-expanding valve was associated with a longer x-ray time (P = .015) and a 2-fold greater first operator radiation dose (P = .018). Patient radiation dose was not significantly affected by arterial approach (P = .055) or valve type (P = .095). After adjustment for potential confounders, the TC approach remained associated with a 174.8% (95% confidence interval, 80.6-318.3, P < .001) increase in first operator radiation dose, whereas the use of a self-expanding valve no longer influenced the first operator dose (P = .630).
Conclusions:
TC access and the use of a self-expanding valve were associated with a 4- and 2-fold greater first operator radiation exposure during TAVR procedures, respectively. Unlike the arterial approach, the effect of bioprosthesis type on radiation exposure was mainly related to x-ray time and was no longer significant after adjustment.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023