Related Experiment Video
Updated: Dec 8, 2025

Ultrasound Cyclo Plasty in Eyes with Glaucoma
Published on: January 26, 2018
Efficacy and safety of high-dose ultrasound cyclo-plasty procedure in refractory glaucoma
He-Ting Liu1, Qing Zhang1, Zheng-Xuan Jiang1
1Department of Ophthalmology, the Second Affiliated Hospital of Anhui Medical University, Hefei 230601, Anhui Province, China.
Aim:
To evaluate the efficacy and safety of high-dose ultrasound cyclo-plasty (UCP) for the treatment of refractory glaucoma in Chinese patients.
Methods:
In this 6-month retrospective study, 37 eyes of 37 patients suffering from severe glaucoma with uncontrolled intraocular pressure (IOP) of ≥21 mm Hg underwent 8-s ultrasonic cyclocoagulation with ten active piezoelectric elements. A complete ophthalmic examination was performed before and at 1d, 1, 3, 6mo after UCP. Therapeutic success was defined as IOP reduction from baseline ≥20% and IOP ≥5 mm Hg without adding new glaucoma medication compare to baseline at the 6-month follow-up visit. In addition to mean IOP at each follow-up visit, medications used and complications were also detected and compared to baseline.
Results:
After UCP procedure, the mean IOP was significantly reduced (P<0.01) from the preoperative 44.1±11.9 mm Hg to postoperative 26.7±11.8 mm Hg at 3mo, and 30.4±14.5 mm Hg at 6mo. The overall mean IOP reductions achieved at 3 and 6mo were 39% and 31% compared to baseline IOP. Sixty-one percent of patients responded well to UCP treatment with a mean IOP reduction of 48% at 3mo and 42% at 6mo. Ocular pain in most of patients were alleviated. No serious intraoperative or postoperative complications occurred.
Conclusion:
High-dose UCP treatment is an effective and safe procedure to reduce IOP in Chinese patients with severe glaucoma.