Comparison between his-bundle pacing and left bundle branch pacing in patients with atrioventricular block
1Department of Cardiology and Macrovascular Disease, Beijing Tiantan Hospital, Capital Medical University, No. 119 South Fourth Ring West Road, Beijing, 100070, China.
Background:
Pacing the cardiac conduction system has been explored in patients with conduction system disease, but comprehensive comparisons between different pacing modalities are not well investigated.
Objective:
To compare pacing characteristics and ventricular synchrony between His-bundle pacing (HBP) and left bundle branch pacing (LBBP) in patients with atrioventricular block (AVB).
Methods:
Fifty pacemaker-indicated patients with AVB were enrolled. Twenty-five patients underwent HBP, and another 25 patients underwent LBBP. Success rate, procedural and fluoroscopy duration, pacing parameters, and echocardiographic data were perioperatively assessed and at 3-month follow-up.
Results:
HBP was successful in 19 of 25 (76.0%) patients, whereas LBBP was successful in 22 of 25 (88.0%) patients. Compared with HBP, LBBP capture threshold was significantly lower (0.76 ± 0.25 V/0.4 ms vs. 1.27 ± 0.61 V/1.0 ms, P = 0.003) and R-wave amplitude was significantly higher with LBBP (11.7 ± 6.6 vs. 4.9 ± 2.4 mV, P < 0.001) at implant. The mean procedural time (74.3 ± 17.8 vs. 63.2 ± 12.3 min, P = 0.029) and fluoroscopy duration (10.3 ± 4.5 vs. 6.8 ± 2.2 min, P = 0.005) were significantly longer in the HBP group compared to LBBP. At 3-month follow-up, pacing capture threshold remained more stable in LBBP than in HBP group while left ventricular synchrony was similar between both groups.
Conclusion:
Despite similar impact on ventricular synchrony compared with HBP, LBBP featured a significantly lower pacing capture threshold, higher R-wave amplitude, and less time to achieve similar success rate in patients with AVB. These findings indicate LBBP as a physiological pacing strategy for AVB patients.
Related Concept Videos
Dysrhythmias IV: Characteristics of Bradyarrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Conduction System of the Heart
This system relies on the unique properties of nodal and Purkinje cells:...
The Cardiac Cycle
The Process
Electrical signals—sent from the sinoatrial (SA) node in the right atrial wall to the atrioventricular (AV) node between the right atrium and right ventricle—cause both atria to simultaneously contract. When the signal reaches the AV node, it pauses for approximately a tenth of a second, allowing the atria to contract and...