Mimicking and Measuring Occlusal Erosive Tooth Wear with the "Rub&Roll" and Non-contact Profilometry
Published on: February 2, 2018
Novel methodology for measuring intraoral wear in enamel and dental restorative materials
Josephine F Esquivel-Upshaw1, Shu-Min Hsu1, Ana C Bohórquez2
1Restorative Dental Sciences, Division of Prosthodontics, University of Florida College of Dentistry, Gainesville, Florida, USA.
Objectives:
To test the hypotheses that (a) the chairside/handheld dental scanner combined with a metrology software will measure clinical wear in vivo in agreement with measurements from X-ray computed microtomography and; (b) polished monolithic zirconia does not cause accelerated wear of opposing enamel.
Materials And Methods:
Thirty single crowns were randomized to receive a monolithic zirconia or metal-ceramic crown. Two non-restored opposing teeth in the same quadrants were identified to serve as enamel controls. After cementation, quadrants were scanned using an intraoral dental scanner. Patients were recalled at 6-months and 1-year for re-scanning. Scanned images were compared using a metrology software to determine maximum vertical wear of teeth. The accuracy of the scanning measurements from this new method was compared with X-ray computed microtomography (micro-CT) measurements. Statistical analysis was performed using Mann-Whitney U test to determine significant differences between wear of enamel against zirconia, metal-ceramic or enamel. Linear regression analysis determined agreement between measurements obtained using intraoral scanning and micro-CT.
Results:
Regression analysis demonstrated that there is a quantitative agreement between depth and volume measurements produced using intraoral scanning and the micro-CT methodologies. There was no significant difference between the wear of enamel against polished monolithic zirconia crowns and enamel against enamel.
Conclusions:
Intraoral scanning combined with a matching software can accurately quantify clinical wear to verify that monolithic zirconia exhibited comparable wear of enamel compared with metal-ceramic crowns and control enamel. Agreement between the intraoral scanner and the micro-CT was 99.8%. Clinical Trials.gov NCT02289781.