Dose reduction potential of using gold fiducial markers for kilovoltage image-guided radiotherapy
Yoshiki Takei1,2, Hajime Monzen1, Mikoto Tamura1
1Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, Osakasayama, Japan.
Abstract:
This study aimed to evaluate the possibility of reducing the imaging dose for image-guided radiotherapy by using planar kilovoltage orthogonal imaging and fiducial markers (kV-FM). We tested kilovoltage planar images under clinical imaging conditions for the pelvis (75 kVp, 200 mA, 50 ms) at a decreasing tube current (from 200 to 10 mA). Imaging doses were measured with a semiconductor detector. The visibility of the kV-FM, aspects of image quality (spatial resolution, low contrast resolution), and the resultant image registration reproducibility were evaluated using various shapes (folded, linear, tadpole-like) of fiducial markers containing 0.5% iron [Gold Anchor™ (GA); Naslund Medical AB, Huddinge, Sweden]. The GA phantom was created by placing these variously shaped GAs in an agar phantom. The imaging doses with 200 and 10 mA were approximately 0.74 and 0.04 mGy and they were correlated to the tube current (R2 = 0.999). Regardless of the marker's shape, the GA phantom ensured visibility even when the tube current was reduced to the minimum value (10 mA). The low contrast resolution was gradually decreased at less than 50 mA, but the spatial resolution did not change. Although the auto-registration function could not be used, manual-registration could be achieved with an accuracy of within 1 mm, even when the imaging dose was reduced to 1/20 of the clinical imaging condition for the pelvis. When using the GA as the fiducial marker, the imaging dose could be reduced to 1/20 of that used clinically while maintaining the accuracy of manual-registration using the kV-FM for image-guided radiotherapy of the pelvis.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014