Temperature Elevation in an Instrumented Phantom Insonated by B-Mode Imaging, Pulse Doppler and Shear Wave
Maha Issaoui1, Piero Miloro2, Xavier Balandraud1
1Université Clermont Auvergne, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France.
Ultrasound in Medicine & Biology
|September 23, 2020
Summary
Shear wave elastography (SWE) measures tissue elasticity in obstetrics but its fetal safety is unknown. This study measured temperature increases from diagnostic ultrasound modes, finding SWE caused the most heating, necessitating further safety evaluations.
Area of Science:
- Medical Imaging
- Biophysics
- Obstetrics and Gynecology
Background:
- Diagnostic ultrasound is crucial for obstetric care, providing anatomical and physiological data.
- Shear wave elastography (SWE) assesses tissue elasticity and is expanding into obstetrics and gynecology.
- The thermal and mechanical effects of SWE on fetal development are currently unknown.
Purpose of the Study:
- To measure the temperature increase caused by B-mode, pulse Doppler (PD), and SWE.
- To provide reference data for future obstetric and gynecologic SWE safety studies.
- To offer initial recommendations on SWE safety in these clinical areas.
Main Methods:
- Experiments were conducted using an Aixplorer diagnostic ultrasound system.
- An instrumented phantom with 11 embedded thermocouples was used to measure temperature changes.
- Temperature increases were recorded for B-mode, PD, and SWE modes after 380-second excitation.
Main Results:
- The highest temperature increase was observed near the transducer surface.
- SWE caused a maximum temperature increase of 2.9°C, compared to 1.2°C for PD and 0.7°C for B-mode.
- Both transducer conduction and ultrasound wave heating contributed to the temperature rise, with SWE showing greater heating.
Conclusions:
- SWE demonstrates a higher potential for thermal effects compared to B-mode and PD ultrasound.
- Further research and regulatory evaluation are essential to determine the safety of SWE in obstetric and gynecologic applications.
- Methodological data is provided to guide future safety assessments for SWE in pregnancy.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
2.1K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
2.1K
Equipments Used To Measure Blood Pressure
2.8K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
2.8K