Comparison of virtual to true unenhanced abdominal computed tomography images acquired using rapid kV-switching dual
D Olivia Popnoe1, Chaan S Ng2, Shouhao Zhou3
1MD Anderson Cancer Center UT Health Graduate School of Biomedical Sciences, Houston, Texas, United States of America.
Objective:
To compare "virtual" unenhanced (VUE) computed tomography (CT) images, reconstructed from rapid kVp-switching dual-energy computed tomography (DECT), to "true" unenhanced CT images (TUE), in clinical abdominal imaging. The ability to replace TUE with VUE images would have many clinical and operational advantages.
Methods:
VUE and TUE images of 60 DECT datasets acquired for standard-of-care CT of pancreatic cancer were retrospectively reviewed and compared, both quantitatively and qualitatively. Comparisons included quantitative evaluation of CT numbers (Hounsfield Units, HU) measured in 8 different tissues, and 6 qualitative image characteristics relevant to abdominal imaging, rated by 3 experienced radiologists. The observed quantitative and qualitative VUE and TUE differences were compared against boundaries of clinically relevant equivalent thresholds to assess their equivalency, using modified paired t-tests and Bayesian hierarchical modeling.
Results:
Quantitatively, in tissues containing high concentrations of calcium or iodine, CT numbers measured in VUE images were significantly different from those in TUE images. CT numbers in VUE images were significantly lower than TUE images when calcium was present (e.g. in the spine, 73.1 HU lower, p < 0.0001); and significantly higher when iodine was present (e.g. in renal cortex, 12.9 HU higher, p < 0.0001). Qualitatively, VUE image ratings showed significantly inferior depiction of liver parenchyma compared to TUE images, and significantly more cortico-medullary differentiation in the kidney.
Conclusions:
Significant differences in VUE images compared to TUE images may limit their application and ability to replace TUE images in diagnostic abdominal CT imaging.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
X-ray Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...