Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
A Comparative Assessment of the Nova Stat Profile Prime Plus® Critical Care Analyzer
Ali Jawa1, Feroza Motara1, Muhammed Moolla1
1Emergency Medicine, University of the Witwatersrand, Johannesburg, ZAF.
Background:
Point-of-care testing (POCT) plays an integral role in the management of acutely ill patients presenting to the emergency department (ED). Due to its rapid turnaround time, POCT has been shown to improve ED workflow, reduce unnecessary admissions and lessen the burden on ED staff. The aim of the study was to compare the accuracy, precision and linearity of the Nova Stat Profile Prime Plus® (Nova Biomedical, Waltham, MA, USA) to the Radiometer ABL800 FLEX® (Radiometer South Africa Pty Ltd, Gauteng) and the Abbott i-stat Chem8+® (Abbott, Princeton, NJ, USA) POCT analyzers.
Methods:
A convenience sample of 150 discarded whole blood specimens was obtained and analyzed. Paired test measurements were conducted for method comparison. Accuracy was measured by pairing individual results from the Nova Stat Profile Prime Plus® with either the Radiometer ABL800 FLEX® or the Abbot i-stat Chem8+® analyzers by calculating the differences.
Results:
The with-in run percentage coefficient of variation (%CV) was below 2.4% for pH, carboxyhemoglobin (COHb), deoxyhemoglobin (HHb), total hemoglobin (tHb), total bilirubin (tBil), sodium (Na), potassium (K), chloride (Cl), ionized calcium (iCa), urea, glucose and lactate, and was below 5.1% for all other analytes. The day-to-day %CV was below 1.6% for pH, COHb, HHb, tHb, tBil, Na, K, Cl, iCa, urea, glucose and lactate, and below 6.10% for all other analytes. The correlation coefficient (r) was 0.351 and ranged from 0.897 to 0.998 for all analytes. The mean bias was minimal for all analytes.
Conclusion:
There was a good correlation between the Nova Stat Profile Prime Plus® and the Radiometer ABL800 FLEX®/Abbott i-STAT Chem8+® POCT analyzers. The Stat Profile Prime Plus® exhibited good precision both within-run and day-to-day.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
06:16Signal Acquisition, Score Interpretation, and Economics of a Non-Invasive Point-of-Care Test for Coronary Artery Disease
Published on: August 9, 2024
Related Concept Videos
Pulse Assessment Sites
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like...
Acute Coronary Syndrome III: Diagnostic Studies
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: