Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Technical Note: Fricke dosimetry for blood irradiators
Andrea Mantuano1, Camila Salata1,2, Carla Lemos Mota3
1Radiological Sciences Department, Rio de Janeiro State University (UERJ), Rio de Janeiro, Brazil.
Purpose:
The Fricke dosimeter has been shown to be a viable option as an absorbed dose standard. This work aims to provide the dose distribution in an irradiator container during blood irradiation using Fricke dosimetry.
Methods:
Measurements were performed using a Gammacell Elan 3000 blood irradiator at Hemocenter in Rio de Janeiro, Brazil. A specific phantom was constructed and patented by the authors to perform these measurements. Fricke solution was prepared according to international protocols, and polyethylene bags filled with Fricke solution (n = 19) were spatially distributed within the phantom. Control bags were also submitted to the same process, except the irradiation. The irradiation time was calculated to give 25.7 Gy to the central portion of the phantom, the same dose used for blood bags.
Results:
Encouraging results were obtained with an overall uncertainty of 2.1% (k = 1). The obtained results were compared with the doses calculated by the physicist from Hemocenter based on parameters provided by the manufacturer. The mean dose delivered to the Fricke bag in the center of the phantom (cavity 2) was 28.7 ± 0.5 Gy, which is 12% higher than the planned dose of 25.7 Gy.
Conclusions:
The obtained results showed that the setup (Fricke and phantom) is able to perform dosimetry for blood irradiators. The delivered dose was higher than expected. This highlights the importance in controlling all the parameters during irradiation to ensure the correct dose for all irradiated bags.