Automated enzymatic assay for plasma ammonia
Clinical Chemistry
|January 1, 1979
Summary
A new enzymatic method accurately measures plasma ammonia using the Du Pont Automatic Clinical Analyzer (ACA). This optimized assay offers high sensitivity and avoids interferences, improving clinical diagnostics.
Area of Science:
- Clinical Chemistry
- Enzymology
- Analytical Biochemistry
Background:
- Accurate plasma ammonia determination is crucial for diagnosing hepatic encephalopathy and other metabolic disorders.
- Existing methods may suffer from interferences or require lengthy pre-incubation steps.
- The Du Pont Automatic Clinical Analyzer (ACA) offers a platform for automated clinical assays.
Purpose of the Study:
- To develop and validate a sensitive enzymatic method for plasma ammonia measurement using the ACA.
- To optimize reaction conditions for enhanced assay performance.
- To assess potential interferences in the developed assay.
Main Methods:
- An enzymatic assay utilizing glutamate dehydrogenase and monitoring NADPH depletion at 340 nm.
- Kinetic measurement performed with a sample volume of 500 muL on the ACA.
- Optimization of reaction conditions using one-factor-at-a-time and response surface methodology.
- Evaluation of interferences from anticoagulants and common endogenous blood components.
Main Results:
- The assay demonstrated linearity up to 1000 mumol/L of plasma ammonia.
- No significant interferences were observed from anticoagulants, pyruvate, or oxalacetate.
- Use of NADPH eliminated the need for pre-incubation, reducing assay time.
- Optimized conditions yielded maximum assay sensitivity.
Conclusions:
- The described enzymatic ACA method provides a sensitive, linear, and interference-free assay for plasma ammonia.
- This method offers a reliable and efficient tool for clinical laboratories.
- The assay's robustness makes it suitable for routine diagnostics.