Stop the leak!: Mitigating potential exposure of aerosolized COVID-19 during laparoscopic surgery
John M Uecker1,2, Austin Fagerberg1, Naser Ahmad1
1ClearCam Inc, Austin, TX, 78744, USA.
Background:
Viral particles have been shown to aerosolize into insufflated gas during laparoscopic surgery. In the operating room, this potentially exposes personnel to aerosolized viruses as well as carcinogens. In light of circumstances surrounding COVID-19 and a concern for the safety of healthcare professionals, our study seeks to quantify the volumes of gas leaked from dynamic interactions between laparoscopic instruments and the trocar port to better understand potential exposure to surgically aerosolized particles.
Methods:
A custom setup was constructed to simulate an insufflated laparoscopic surgical cavity. Two surgical instrument use scenarios were examined to observe and quantify opportunities for insufflation gas leakage. Both scenarios considered multiple configurations of instrument and trocar port sizes/dimensions: (1) the full insertion and full removal of a laparoscopic instrument from the port and (2) the movement of the scope within the port, recognized as "dynamic interaction", which occurs nearly 100% of the time over the course of any procedure.
Results:
For a 5 mm instrument in a 5 mm trocar, the average volume of gas leaked during dynamic interaction and full insertion/removal scenarios were 43.67 and 25.97 mL of gas, respectively. Volume of gas leaked for a 5 mm instrument in a 12 mm port averaged 41.32 mL and 29.47 for dynamic interaction vs. instrument insertion and removal. Similar patterns were shown with a 10 mm instrument in 12 mm port, with 55.68 mL for the dynamic interaction and 58.59 for the instrument insertion/removal.
Conclusions:
Dynamic interactions and insertion/removal events between laparoscopic instruments and ports appear to contribute to consistent leakage of insufflated gas into the OR. Any measures possible taken to reduce OR gas leakage should be considered in light of the current COVID-19 pandemic. Minimizing laparoscope and instrument removal and replacement would be one strategy to mitigate gas leakage during laparoscopic surgery.
Related Concept Videos
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Suctioning the Nasopharyngeal Airway
Equipment Required