Lifecycle evidence requirements for high-risk implantable medical devices: a European perspective
Rosanna Tarricone1,2,3, Oriana Ciani3,4, Aleksandra Torbica1,2
1Department of Social and Political Science, Bocconi University , Milan, Italy.
Introduction:
The new European Union (EU) Regulations on medical devices and on in vitro diagnostics provide manufacturers and Notified Bodies with new tools to improve pre-market and post-market clinical evidence generation especially for high-risk products but fail to indicate what type of clinical evidence is appropriate at each stage of the whole lifecycle of medical devices. In this paper we address: i) the appropriate level and timing of clinical evidence throughout the lifecycle of high-risk implantable medical devices; and ii) how the clinical evidence generation ecosystem could be adapted to optimize patient access.
Areas Covered:
The European regulatory and health technology assessment (HTA) contexts are reviewed, in relation to the lifecycle of high-risk medical devices and clinical evidence generation recommended by international network or endorsed by regulatory and HTA agencies in different jurisdictions.
Expert Opinion:
Four stages are relevant for clinical evidence generation: i) pre-clinical, pre-market; ii) clinical, pre-market; iii) diffusion, post-market; and iv) obsolescence & replacement, post-market. Each stage has its own evaluation needs and specific studies are recommended to generate the appropriate evidence. Effective lifecycle planning requires anticipation of what evidence will be needed at each stage.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
08:17Autonomous and Rechargeable Microneurostimulator Endoscopically Implantable into the Submucosa
Published on: September 27, 2018
Related Concept Videos
Bacterial Signaling
Clinically Relevant Drug Product Specifications: Methods of Establishment
Drug Products: Biologics, Biosimilars and Interchangeables