Evaluation of Spatial Resolution of Virtual Monochromatic Imaging In Vitro: Effect of Energy Level and Contrast
Shigeru Suzuki1, Rika Fukui1, Yuzo Yamamoto1
1From the Department of Radiology, Tokyo Women's Medical University Medical Center East.
Objective:
Our aim was to evaluate effects of the energy level and contrast on the spatial resolution of virtual monochromatic imaging (VMI).
Methods:
With 2 types of computed tomographic (CT) scanners (Discovery CT750 HD and Revolution CT), we scanned an elliptical cylinder acrylic phantom with 20- or 40-fold dilutions of contrast medium and obtained VMI data sets at 40, 50, 60, and 70 keV. We evaluated the effects of energy levels and contrast on modulation transfer function (MTF) and the effect of energy levels on noise power spectra (NPS).
Results:
With both CT scanners, MTF decreased significantly as the energy level decreased. The effect of the dilution of contrast media on MTF varied with CT scanners. With both scanners, NPS curves demonstrate uniform increase in NPS across the entire spatial frequency as the energy levels decreased.
Conclusions:
The energy level and contrast can affect the spatial resolution of VMIs.
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Super-resolution Fluorescence Microscopy
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Overview of Electron Microscopy
Three-Dimensional Microscopy in Microbiology