Reflections on the Ultrasound Mirror Image Artifact
Dana Middleton Kerr1, William Dana Middleton2
1Department of Radiology, Duke University, Durham, NC.
Abstract:
A key assumption in ultrasound is that the reflected echoes that return to the transducer travel in a straight line. Mirror image artifacts occur when the transmitted pulse and returning echo reflect off of a highly reflective interface (an acoustic mirror) and change direction before returning to the transducer, thereby breaking this assumption. Mirror image artifacts are seen throughout the body on gray scale, color Doppler, power Doppler, and spectral Doppler. They may closely duplicate the true structure in shape and echo strength, may distort the true structure, may appear weaker than the true structure, or may appear on images that do not simultaneously show the true structure. If not properly recognized, mirror image artifacts can be misinterpreted as true pathology and lead to additional unnecessary tests and potentially harmful interventions.
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies II: Ultrasonography
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...