Related Experiment Video
Updated: Dec 7, 2025

Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level
Published on: June 6, 2018
Application of a liquid scintillation system with 100-ml counting vials for environmental tritium determination:
Bin Feng1, Bo Chen1, Chao Zhao2
1Institute of Radiation Medicine, Fudan University, Shanghai, 200032, China.
Abstract:
For more efficient and accurate determination of airborne tritium in the environment, the procedure optimization, performance test and uncertainty analysis of a commercially available low-background liquid scintillation counting (LSC) system with 100-ml counting vials were studied in this work. The results showed that 50 ml water sample mixed with 50 ml scintillation cocktail (Ultima Gold uLLT, PE) could achieve the optimal counting condition after a dark adaption time longer than 1440 min. The minimum detectable activity (MDA) of the 100-ml vial system was estimated to be 0.18 Bq·L-1 in a continuous counting time of 3600 min, which was approximately 3.5 times lower than that of 20-ml vial system, and its determination uncertainty was also generally lower provided the collected samples was more than 15 ml. It indicates that the LSC system with 100-ml counting vials is preferable for environmental tritium determination. However, for more accurate determination, the electrolytic enrichment is still needed for the sample with the specific activity lower than 0.4 Bq·L-1. On the other hand, considering the cost and potential environmental impact of present available cocktails, the system with 20-ml vials is recommended for determining the sample with the specific activity higher than 2 Bq·L-1.
More Related Videos
10:52Radiolabeling and Quantification of Cellular Levels of Phosphoinositides by High Performance Liquid Chromatography-coupled Flow Scintillation
Published on: January 6, 2016
09:58A Modular Microfluidic Technology for Systematic Studies of Colloidal Semiconductor Nanocrystals
Published on: May 10, 2018