Introduction:
A high-performance liquid chromatographic-diode array detector (HPLC-DAD) method for the determination of amoxicillin in medicated feedingstuffs was developed and validated. The method was used to investigate the quality requirements of animal feedingstuffs (declared content of active substance and feed homogeneity).
Material And Methods:
Two-gram samples were extracted by potassium phosphate buffer solution. Extracts were filtered and directly analysed by HPLC-DAD without further clean-up. Amoxicillin was separated by acetonitrile and 0.01M phosphate buffer (pH 5.0) on a Phenomenex Luna C18 column.
Results:
This method provided average recoveries of 76.1 to 81.6% with coefficients of variation (CV, %) for repeatability and reproducibility in the ranges of 3.7-7.2% and 5.3-7.6%, respectively. The limit of detection was 51.2 mg/kg and limit of quantification was 103.0 mg/kg.
Conclusion:
The method was successfully validated and proved to be efficient, precise, and useful for quantification of amoxicillin in medicated feedingstuffs.