Outcomes of single‑lead VDD pacemakers in atrioventricular blocks: The OSCAR study
Giuseppe D Sanna1, Giuseppe Nusdeo1, Alessandro Marini1
1Clinical and Interventional Cardiology, Sassari University Hospital, Sassari, Italy.
Background:
VDD pacemakers are regarded as a second choice in patients with atrio-ventricular blocks mainly due to the potential failure of atrial sensing, leading to a loss of atrio-ventricular synchrony. This single-centre study aimed to evaluate the prevalence of loss of atrial sensing and its potential determinants in patients with VDD pacemakers.
Method:
142 patients with an implanted VDD device underwent long-term follow-up with clinical evaluation, electrocardiogram, device interrogation and echocardiogram.
Results:
Over a long follow-up period [median 110 (68-156) months], 17 patients (12%) in sinus rhythm presented loss of atrial sensing. This was most often intermittent, but three patients required a permanent switch to VVI mode. ECG showed higher prevalence of interatrial blocks (50% vs 26.6%, p = 0.057) and longer P wave duration (116 ± 19 vs 105 ± 15 ms, p = 0.019) in patients with loss of atrial sensing. Echocardiography revealed larger left atrial (LA) volumes (p < 0.05) in patients with loss of atrial sensing, and lower LA ejection fraction (0.40 vs 0.47, p = 0.0037) and expansion index (0.63 ± 0.26 vs 0.90 ± 0.31, p = 0.003). P wave duration on ECG proved to be independently associated with loss of atrial sensing on multivariable analysis (OR 1.062, 95% CI 1.015-1.110; p = 0.008). The prevalence of atrial fibrillation and subsequent switch to VVI mode was high (16%).
Conclusions:
In the long-term follow-up, the loss of atrial sensing is present in 12% of patients with implanted VDD pacemakers. ECG and echocardiographic parameters may serve as screening tools for the detection of atrial myopathy which is associated with the loss of atrial sensing.
More Related Videos
08:05Optimization of Transesophageal Atrial Pacing to Assess Atrial Fibrillation Susceptibility in Mice
Published on: June 29, 2022
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Acute Coronary Syndrome III: Diagnostic Studies
Antiarrhythmic Drugs: Class II Agents as β-Adrenergic Blockers
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Antiarrhythmic Drugs: Class III Agents as Potassium Channel Blockers
Depolarizing Blockers: Mechanism of Action
Succinylcholine is the most commonly used depolarizing blocker. Chemically, it constitutes two molecules of acetylcholine joined together by an acetate methyl group. They act on the receptors in the same way as acetylcholine. Because...